In Vivo Measurement of Cerebral SPIO Concentration in Nonhuman Primate Using Magnetic Particle Imaging Detector
Hui, Hui1,3,9; Liu, Jiaojiao8; Zhang, Hui6,7; Zhong, Jing5; He, Jie6,7; Zhang, Bo6,7; Zhang, Haoran6,7; Li, Qin2; Li, Hongjun8; Tian, Jie1,3,4,6,7
刊名IEEE MAGNETICS LETTERS
2023
卷号14页码:5
关键词Magnetic instruments cerebral blood volume magnetic particle imaging nonhuman primate superparamagnetic iron oxide nanoparticles
ISSN号1949-307X
DOI10.1109/LMAG.2023.3281933
通讯作者Li, Qin(liqin2@motac.com) ; Li, Hongjun(lihongjun00113@ccmu.edu.cn) ; Tian, Jie(tian@ieee.org)
英文摘要The purpose of this study is to develop a magnetic particle imaging (MPI) technique to directly measure time-varied cerebral superparamagnetic iron oxide nanoparticle (SPIO) concentration in rhesus macaques. A hand-held MPI detector was developed to monitor MPI signal changes at the third harmonics of the drive frequency in resting-state nonhuman primates. Phantom experiments were first performed to determine the sensitivity limits of the detector as a function of distance from the detector and SPIO concentration. The measured sensitivity profile was then used to reveal the most sensitive region of the detector. MPI detection was continuously performed to monitor MPI signal changes after two bolus injections of SPIOs in the rhesus macaque. We successfully developed a hand-held MPI to detect cerebral SPIO concentration changes in a living nonhuman primate. The detection limit of the MPI detector is about 125 ng iron. We reported on the in vivo measurement of cerebral SPIO concentration changes in rhesus macaque using a hand-held MPI detector. In vivo experiments showed the feasibility of the detector to sensitively measure MPI signals in a nonhuman primate brain.
资助项目Beijing Natural Science Foundation[JQ22023] ; Beijing Natural Science Foundation[7212051] ; Beijing Natural Science Foundation[L222097] ; National Key Research and Development Program of China[2017YFA0700401] ; National Key Research and Development Program of China[Z211100003521003] ; National Natural Science Foundation of China[62027901] ; National Natural Science Foundation of China[81827808] ; National Natural Science Foundation of China[61936013] ; National Natural Science Foundation of China[81771806] ; National Natural Science Foundation of China[81227901] ; CAS Youth Innovation Promotion Association[Y2022055] ; CAS Key Technology Talent Program
WOS关键词BLOOD-VOLUME ; CONTRAST AGENT ; FMRI ; TOMOGRAPHY ; TRACERS ; MOTION ; AWAKE ; BOLD
WOS研究方向Engineering ; Physics
语种英语
出版者IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
WOS记录号WOS:001025460700001
资助机构Beijing Natural Science Foundation ; National Key Research and Development Program of China ; National Natural Science Foundation of China ; CAS Youth Innovation Promotion Association ; CAS Key Technology Talent Program
内容类型期刊论文
源URL[http://ir.ia.ac.cn/handle/173211/53861]  
专题自动化研究所_中国科学院分子影像重点实验室
通讯作者Li, Qin; Li, Hongjun; Tian, Jie
作者单位1.Beijing Key Lab Mol Imaging, Beijing 100190, Peoples R China
2.Motac Beijing Serv Ltd, Beijing 100050, Peoples R China
3.Chinese Acad Sci, Inst Automat, Key Lab Mol Imaging, Beijing 100190, Peoples R China
4.Beihang Univ, Key Lab Big Data Based Precis Med, Minist Ind & Informat Technol China, Beijing 100191, Peoples R China
5.Beihang Univ, Sch Instrumentat & Optoelect Engn, Beijing 100191, Peoples R China
6.Beihang Univ, Sch Biol Sci & Med Engn, Beijing 100191, Peoples R China
7.Beihang Univ, Sch Engn Med, Beijing 100191, Peoples R China
8.Capital Med Univ, Beijing Youan Hosp, Dept Radiol, Beijing 100069, Peoples R China
9.Univ Chinese Acad Sci, Beijing 100080, Peoples R China
推荐引用方式
GB/T 7714
Hui, Hui,Liu, Jiaojiao,Zhang, Hui,et al. In Vivo Measurement of Cerebral SPIO Concentration in Nonhuman Primate Using Magnetic Particle Imaging Detector[J]. IEEE MAGNETICS LETTERS,2023,14:5.
APA Hui, Hui.,Liu, Jiaojiao.,Zhang, Hui.,Zhong, Jing.,He, Jie.,...&Tian, Jie.(2023).In Vivo Measurement of Cerebral SPIO Concentration in Nonhuman Primate Using Magnetic Particle Imaging Detector.IEEE MAGNETICS LETTERS,14,5.
MLA Hui, Hui,et al."In Vivo Measurement of Cerebral SPIO Concentration in Nonhuman Primate Using Magnetic Particle Imaging Detector".IEEE MAGNETICS LETTERS 14(2023):5.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace