International comparison of the SIAR measurement and the WRR standard
X.Tang; Y.-z.Xia; W.Fang; Y.-p.Wang; X.Ye
刊名Optoelectronics Letters
2019
卷号15期号:2页码:147-150
关键词Uncertainty analysis,Budget control,Radiometers,Solar radiation
ISSN号16731905
DOI10.1007/s11801-019-8117-2
英文摘要We experimentally compare the solar irradiance absolute radiometer (SIAR) measurement with the world radiometric reference (WRR) standard to improve the accuracy of instrument. The SIAR joined in the international pyrheliometer comparison (IPC) in 2000. The comparison results show that the calibration factors for SIAR to WRR are 0.999 220, 1.001 694, 0.998 334 and 0.997 439 in the 9th IPC, the 10th IPC, the 11th IPC and the 12th IPC, respectively. These results are added to the measurement uncertainty budget of SIAR. The repeatability of the SIAR-type absolute radiometers is also investigated. The relative error introduced by two SIAR-type absolute radiometers is within 0.25%. 2019, Tianjin University of Technology and Springer-Verlag GmbH Germany, part of Springer Nature.
URL标识查看原文
内容类型期刊论文
源URL[http://ir.ciomp.ac.cn/handle/181722/63048]  
专题中国科学院长春光学精密机械与物理研究所
推荐引用方式
GB/T 7714
X.Tang,Y.-z.Xia,W.Fang,et al. International comparison of the SIAR measurement and the WRR standard[J]. Optoelectronics Letters,2019,15(2):147-150.
APA X.Tang,Y.-z.Xia,W.Fang,Y.-p.Wang,&X.Ye.(2019).International comparison of the SIAR measurement and the WRR standard.Optoelectronics Letters,15(2),147-150.
MLA X.Tang,et al."International comparison of the SIAR measurement and the WRR standard".Optoelectronics Letters 15.2(2019):147-150.