Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM
Zhang, Dali; Li, Xinqiao; Wen, Xiangyang; Xiong, Shaolin; An, Zhenghua; Sun, Xilei; Qiao, Rui; Li, Zhengwei; Gong, Ke; Hou, Dongjie
学科主题Physics
关键词Instrumentation and Detectors High Energy Physics - Experiment
英文摘要Each satellite of the Gravitational wave high-energy Electromagnetic Counterpart All-sky Monitor (GECAM, mission) consists of 25 SiPM based gamma-ray detectors (GRDs). Although SiPM based GRD has merits of compact size and low bias-voltage, the drift of the SiPM gain with temperature is a severe problem for GRD performance. An adaptive voltage supply source was designed to automatically adjust the SiPM bias voltage to compensate the temperature effects and keep the gain stable. This approach has been proved to be effective during both the on-ground and in-flight tests. The in-flight measured variation of the SiPM gain is within 2%. To reduce the gain non-uniformity of GRDs, an iterative bias voltage adjustment approach is proposed and implemented. The gain non-uniformity is reduced from 17% to 0.6%. In this paper, the gain stabilization and consistency correction approach are presented and discussed in detail.
2021
语种英语
内容类型预印本
源URL[http://ir.ihep.ac.cn/handle/311005/289906]  
专题引力波暴高能电磁对应体全天监测器(GECAM)
高能物理研究所_实验物理中心
高能物理研究所_粒子天体物理中心
高能物理研究所_多学科研究中心
高能物理研究所_管理与技术支持
高能物理研究所_东莞分部
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Zhang, Dali,Li, Xinqiao,Wen, Xiangyang,et al. Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM. 2021.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace