Determination of total selenium in water by hydride generation atomic fluorescence spectrometry | |
Luo Kun-Li | |
2012 | |
关键词 | Atoms Fluorescence spectroscopy Hydrides Hydrochloric acid Selenium Standards |
英文摘要 | In order to establish a rapid and accurate hydride generation-atomic fluorescence spectrometry method for the determination of selenium content in water samples, the influence factors on Se nitrolysis in the water such as nitrolysis temperature and the amount of hydrochloric acid were studied. The authors determined a simple and practical test method that is to add 3 mL hydrochloric acid at 3C. We measured the precision and accuracy of the method by the standard liquid of 3, 6 and 10 g·L-1, and the test results were 2.96, 5.43 and 9.66 mg·L-1. The recovery varied from 90.50% to 98.67%. Meanwhile we tested 29 water samples by different methods. The error of standard sample was 8.95%~25.46%by hydride generation-atomic fluoresence spectrometry in the GB/T 5750.6-2006, and the error of standard sample was 1.33%~3.40% by new method. Compared with the national standards of GB/T 5750.6-2006, the method has the advantages of operating simplicity and and easy control, and it is also reliable and economical. |
出处 | Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis |
卷 | 32期:2页:532-536 |
收录类别 | EI |