Performance tests for ray-scan 64 PET/CT based on NEMA NU-2 2007
Li, Suying ; Zhou, Kun ; Zhang, Qiushi ; Zhang, Jinming ; Yang, Kun ; Xu, Baixuan ; Ren, Qiushi
刊名JOURNAL OF INSTRUMENTATION
2015
关键词Gamma camera, SPECT, PET PET/CT, coronary CT angiography (CTA) Inspection with gamma rays Image reconstruction in medical imaging Detector design and construction technologies and materials
DOI10.1088/1748-0221/10/03/P03016
英文摘要This paper focuses on evaluating the performance of the Ray-Scan 64 PET/CT system, a newly developed PET/CT in China. It combines a 64 slice helical CT scanner with a high resolution PET scanner based on BGO crystals assembled in 36 rings. The energy window is 350 similar to 650 keV, and the coincidence window is set at 12 ns in both 2D and 3D mode. The transaxial field of view (FOV) is 600 mm in diameter, and the axial FOV is 163 mm. Method: Performance measurements were conducted focusing on PET scanners based on NEMA NU-2 2007 standard. We reported the full characterization (spatial resolution, sensitivity, count rate performance, scatter fraction, accuracy of correction, and image quality) in both 2D and 3D mode. In addition, the clinical images from two patients of different types of tumor were presented to further demonstrate this PET/CT system performance in clinical application. Results: using the NEMA NU-2 2007 standard, the main results: (1) the transaxial resolution at 1cm from the gantry center for 2D and 3D was both 4.5mm (FWHM), and at 10cm from the gantry center, the radial (tangential) resolution were 5.6mm (5.3mm) and 5.4mm (5.2mm) in 2D and 3D mode respectively. The axial resolution at 1cm and 10cm off axis was 3.4mm (4.8mm) and 5.5mm (5.8mm) in 2D (3D) mode respectively; (2) the sensitivity for the radial position R0(r=0mm) and R100(r=100mm) were 1.741 kcps/MBq and 1.767 kcps/MBq respectively in 2D mode and 7.157 kcps/MBq and 7.513 kcps/MBq in 3D mode; (3) the scatter fraction was calculated as 18.36% and 42.92% in 2D and 3D mode, respectively; (4) contrast of hot spheres in the image quality phantom in 2D mode was 50.33% (52.87%), 33.34% (40.86%), 20.64% (26.36%), and 10.99% (15.82%), respectively, in N=4 (N=8). Besides, in clinical study, the diameter of lymph tumor was about 2.4 cm, and the diameter of lung cancer was 4.2 cm. This PET/CT system can distinguish the position of cancer easily. Conclusion: The results show that the performance of the newly developed PET/CT system is of high resolution, and low scatter characteristics, and is suitable for clinical applications.; Special High-performance Medical Diagnostic Equipment Program from Development and Reform Commission in China; National Key Instrumentation Development Project Foundation of China [2011YQ030114]; Guangdong Innovative Research Team Program [2011S090]; National Basic Research Program of China (973 Program) [2011CB707500]; SCI(E); EI; ARTICLE; renqsh@coe.pku.edu.cn; 3; 10
语种英语
内容类型期刊论文
源URL[http://ir.pku.edu.cn/handle/20.500.11897/420714]  
专题工学院
推荐引用方式
GB/T 7714
Li, Suying,Zhou, Kun,Zhang, Qiushi,et al. Performance tests for ray-scan 64 PET/CT based on NEMA NU-2 2007[J]. JOURNAL OF INSTRUMENTATION,2015.
APA Li, Suying.,Zhou, Kun.,Zhang, Qiushi.,Zhang, Jinming.,Yang, Kun.,...&Ren, Qiushi.(2015).Performance tests for ray-scan 64 PET/CT based on NEMA NU-2 2007.JOURNAL OF INSTRUMENTATION.
MLA Li, Suying,et al."Performance tests for ray-scan 64 PET/CT based on NEMA NU-2 2007".JOURNAL OF INSTRUMENTATION (2015).
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace