CORC  > 北京大学  > 信息科学技术学院
Laser pumped He-4 magnetometer with light shift suppression
Lin, Zaisheng ; Wang, He ; Peng, Xiang ; Wu, Teng ; Guo, Hong
2016
关键词POLARIZATION SPECTROSCOPY FREQUENCY-MODULATION DISCHARGE
英文摘要We report a laser-pumped He-4 atomic magnetometer with light shift suppression through the atomic sensor itself. A linearly polarized light is used to optically align the He-4 metastable atoms and we monitor the magneto-optical double resonance (MODR) signals produced by the left-and right-circularly orthogonal components. It is shown that light shift leads to the atomic alignment to orientation conversion effect, and thus, the difference between the two MODR signals. One of these two MODR signals is locked at the Larmor frequency and is used to measure the ambient magnetic field, while the differential signal is, simultaneously, fed back to suppress the light shift. The scheme could be of the advantage to the design of compact magnetometers by reducing the systematic errors due to light shift. Published by AIP Publishing.; CPCI-S(ISTP); 11; 87
语种英语
出处21st Topical Conference on High-Temperature Plasma Diagnostics
DOI标识10.1063/1.4967304
内容类型其他
源URL[http://ir.pku.edu.cn/handle/20.500.11897/459423]  
专题信息科学技术学院
推荐引用方式
GB/T 7714
Lin, Zaisheng,Wang, He,Peng, Xiang,et al. Laser pumped He-4 magnetometer with light shift suppression. 2016-01-01.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace