颅颌运动仿真系统(牙合)接触精度初探
陈磊 ; 张豪 ; 冯海兰 ; 张驰 ; 张凤军
刊名中华口腔医学杂志
2010
卷号45期号:2页码:98-101
关键词计算机 模拟 下颌运动 (牙合)接触 精度 Computers analog Mandibular movement Ocelusal contact Accuracy 运动仿真系统 牙合 接触精度 movement simulation reference model 侧方运动 average distance 下颌牙列 下颌模型 模拟精度 practical application mean value 对照组 theoretical basis 运动终点 系统模拟 平均距离 仿真系统的应用 数字化 试验组Computers
ISSN号1002-0098
其他题名the accuracy of occlusal simulation of virtual craniofacial movement simulation system
中文摘要目的 探讨颅颌运动仿真系统的骀接触模拟精度,为该仿真系统的应用提供依据.方法 制作10副石膏模型并上(牙合)架.用(牙合)架模拟侧方运动,三维扫描侧方运动终点(牙合)架上的上下颌模型,重建数字化上下颌模型作为对照组.运用仿真系统模拟耠架的侧方运动,以仿真系统输出的侧方运动终点的数字化上下颌模型为试验组.通过比较试验组与对照组下颌牙列之间的位置差异评价仿真系统的(牙合)接触模拟精度.结果 仿真系统模拟的下颌牙列与对照组下颌牙列之间的绝对平均距离为(0.18±0.05)mm;在前后左右四个分区中,两组右后牙区之间的绝对平均距离最大,为(0.19±0.07)mm.结论 该仿真系统的体外胎接触模拟精度为0.19mm.
学科主题Dentistry ; Oral Surgery & Medicine
语种中文
公开日期2011-05-23
附注Objective To evaluate the accuracy of occlusal contacts simulated by virtual craniofacial movement simulation system, and to provide theoretical basis for its application. Methods Ten pairs of plaster dental casts were fabricated, and mounted on Kavo PROTA 7 articulator. Lateral movement was produced on the articulator. Then each pair of the casts on the articulator was scanned three dimensionally and used as the reference model. The movement of the articulator was simulated using the virtual craniofacial movement simulation system. Then, the digital casts were established on the simulation system and used as the test model. To evaluate the accuracy of the simulation system, positional differences between the simulated and the reference lower dentition were compared. Results The mean value of the absolute average distance between the simulated and reference lower dentition was (0.18±0.05) mm.Among the left, right, anterior and posterior areas of the dentition, the maximum mean value of the absolute average distance between the simulated and the reference model was found in the right posterior area, which was (0.19± 0.07) mm. Conclusions The in vitro accuracy of occlusal contacts simulated by virtual craniofacial movement simulation system is 0.19 mm, and this could be a reference data for the practical application of this system.
内容类型期刊论文
源URL[http://124.16.136.157/handle/311060/9986]  
专题软件研究所_人机交互技术与智能信息处理实验室_期刊论文
推荐引用方式
GB/T 7714
陈磊,张豪,冯海兰,等. 颅颌运动仿真系统(牙合)接触精度初探[J]. 中华口腔医学杂志,2010,45(2):98-101.
APA 陈磊,张豪,冯海兰,张驰,&张凤军.(2010).颅颌运动仿真系统(牙合)接触精度初探.中华口腔医学杂志,45(2),98-101.
MLA 陈磊,et al."颅颌运动仿真系统(牙合)接触精度初探".中华口腔医学杂志 45.2(2010):98-101.
个性服务
查看访问统计
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。


©版权所有 ©2017 CSpace - Powered by CSpace