气液平衡数据测定装置可靠性的检验 | |
胡兴兰 ; 王金梅 ; 关荐伊 ; 周荣琪 ; HU Xing-lan ; WANG Jin-mei ; GUAN Jian-yi ; ZHOU Rong-qi | |
2010-06-10 ; 2010-06-10 | |
关键词 | 气液平衡 测定装置 可靠性 乙醇-水 水-醋酸 热力学一致性 VLE apparatus reliability ethanol-water acetic acid-water thermodynamic consistency TQ013.1 |
其他题名 | A Reliability Test for The Vapor-Liquid Equilibrium Apparatus |
中文摘要 | 介绍了气液平衡数据测定装置可靠性的一种检验方法。以气相单循环气液平衡仪为实验装置,分别对乙醇-水体系及水-醋酸体系的常压气液平衡(VLE)数据进行了测定,用气相色谱仪分析气、液相组成并与文献值比较,同时用Herington面积法对所测VLE数据进行热力学一致性的检验,从而考察了本实验装置测定气液平衡数据的稳定性及可靠性。; A reliability test for the vapor-liquid equilibrium(VLE) apparatus was presented.The atmospheric VLE data of both the ethanol-water system and the acetic acid-water system were measured respectively with a VLE apparatus and gas chromatograph.Comparing the measured VLE data with those of the literatures and checking the thermodynamic consistency by Herington method,the reliability of the vapor-liquid equilibrium apparatus was then successfully tested.; 河北省科技攻关项目(06213030) |
语种 | 中文 ; 中文 |
内容类型 | 期刊论文 |
源URL | [http://hdl.handle.net/123456789/64046] |
专题 | 清华大学 |
推荐引用方式 GB/T 7714 | 胡兴兰,王金梅,关荐伊,等. 气液平衡数据测定装置可靠性的检验[J],2010, 2010. |
APA | 胡兴兰.,王金梅.,关荐伊.,周荣琪.,HU Xing-lan.,...&ZHOU Rong-qi.(2010).气液平衡数据测定装置可靠性的检验.. |
MLA | 胡兴兰,et al."气液平衡数据测定装置可靠性的检验".(2010). |